A Medtronic Inc. device used to stimulate the brain in the treatment of neurological conditions has a potentially life-threatening flaw that prompted U.S. regulators to issue its highest-level warning, the Star Tribune reports.

The wires on the ends of Medtronic’s deep-brain stimulators can be damaged by a cap designed to protect the delicate electronics while they are being inserted, the Fridley-based company said Thursday in a news release. The wires can be harmed when the connections inside the cap are twisted during surgery or when the cap is removed, the company said.

The Star Tribune reports that the U.S. Food and Drug Administration classified the issue as a Class I recall, meaning there is a reasonable probability that the defect could cause serious health consequences or death.

Medtronic alerted physicians to the problem in February and has changed the manufacturing in an effort to fix the issue, a change that is under review by the FDA, the Star Tribune reports.

The company urges patients with questions to talk to their doctors.

Next Up

target

Target profits plunge as it takes hit from shedding excess goods

The retailer announced earlier this year it would be slashing prices.

police lights

Las Vegas man identified as victim killed near 38th and Chicago

The 25-year-old died at the scene Sunday afternoon.

unnamed-5

Beloved Eagan Dairy Queen says goodbye

A summer staple has shuttered in Eagan.

Screen Shot 2022-08-16 at 3.44.22 PM (2)

Bridge damage closes westbound Hwy. 62 between Richfield, Edina

The Minnesota Department of Transportation said bridge maintenance work is in progress.

Michael Miller

Charges: Drunk driver killed former chair of Hibbing Chamber of Commerce

Michael Miller's blood-alcohol content was nearly three times the legal limit at the time of the crash.

Mystic Lake

Charges: Man attacked Mystic Lake security with 2x4

A chaotic scene unfolded after the man approached the bell stand late Friday.

Related

Medtronic

Medtronic issues alert for vulnerable defibrillators, home devices

The Department of Homeland Security is aware of the issue.